Trace:
Differences
This shows you the differences between two versions of the page.
Both sides previous revision Previous revision | |||
gp_simulator:usrp_calibration_procedure [2019/02/27 10:22] yasko |
gp_simulator:usrp_calibration_procedure [2019/02/27 15:06] (current) admin |
||
---|---|---|---|
Line 11: | Line 11: | ||
\\ | \\ | ||
- | * Connect RF outlet of the USRP to the spectrum analyzer or RF power meter. When using the spectrum analyzer, we recommend setting its **CF** to **1575.42 MHz**, **bandwidth** to **10 MHz**, and switching on the power measurement mode; | + | * Connect RF port of the USRP to the spectrum analyzer or RF power meter. When using the spectrum analyzer, we recommend set **CF** to **1575.42 MHz**, **bandwidth** to **10 MHz**, and switching on the power measurement mode; |
* Set "**DUT input power, dBm**" to "**-20 dBm**" respectively; | * Set "**DUT input power, dBm**" to "**-20 dBm**" respectively; | ||
- | * Click on the "Start Scenario" button to finish; | + | * Click on the "Start Scenario" button to finish; |
{{:gp_simulator:gp-simulator_usrp_start_calibration.png?600|}} | {{:gp_simulator:gp-simulator_usrp_start_calibration.png?600|}} |